SPG SERIES SINGLE POINT LOAD CELLS, C6 CLASS, from 7 to 36kg

CAPACITIES (kg): 7, 18, 36.

Dit product is momenteel niet op voorraad.

Artikelnummer: Niet beschikbaar Categorie:

Beschrijving

SPG SERIES SINGLE POINT LOAD CELLS, C6 CLASS

Load cells for off center loads, suitable for example to build CE-M
approved scales with a single load cell; maximum platform dimensions of 450×450 mm.
CAPACITIES (kg): 7, 18, 36.
OIML R60. C6 precision class.
Aluminium alloy. IP67 protection.

TECHNICAL FEATURES

Minimum load cell verification interval: Vmin = EMax / 20.000
Maximum number of verification intervals: nLC = 6000
Sensitivity: 2mV/V ± 10%
Combined error: 0,03% of Full Span (F.S.)
Temperature effect on zero: 0,01% FS/10K
Temperature effect on Full Span: 0,001% FS/10K
Compensated Temperature Range: -10°C/+40°C
Operating Temperature Range: -10°C/+50°C
Creep at nominal load in 30 minutes: 0,01% F.S.
Maximum excitation voltage without damage: 15 Vdc
Input Resistance: 300..500 Ohm
Output Resistance: 300..500 Ohm
Zero Balance: 0 ± 0,1% F.S.
Insulation Resistance: >2000 MOhm
Mechanical limit load in service: 150% F.S.
Destructive load: 300% F.S.
Deflection at nominal load <0,3mm
6-wire shielded cable, 3m long, 4,7mm diameter

ATEX Version

The CCATEX certificate makes the load cell suitable for use in
hazardous areas with protection methods according to:
ATEX II 1G Ex ia IIC T6 (Ta -20÷+40°C) TX (Ta -20÷+65°C) Ga
ATEX II 1D Ex ta IIIC TX°C (Ta -20÷+40°C) TX°C (Ta -20÷+65°C)
Da IP65. The certificate includes the manual and the ATEX CE
conformity declaration (available in EN, DE, FR, and IT).
IMPORTANT: the certificate refers to the single load cell.
For systems with more load cells, it is necessary to order one certificate for each load cell.

If you want to receive more information or an offfer you can always contact us.

Average Rating
0 out of 5 stars. 0 votes.

Beoordelingen

Er zijn nog geen beoordelingen.

Wees de eerste om “SPG SERIES SINGLE POINT LOAD CELLS, C6 CLASS, from 7 to 36kg” te beoordelen

Het e-mailadres wordt niet gepubliceerd. Vereiste velden zijn gemarkeerd met *